Stirred Fluid Loss Tester
#120-70 - 115-Volt
#120-70-1 - 230-Volt
Successfully cementing the casing string of
an oil or gas well is highly dependent upon the
characteristics of the cement slurry. Properties
that should be considered include consistency,
density, the ability to quickly develop compressive
strength, rheological properties, and filtration
control. Well cements that have poor filtration
control can lead to a complete failure of the
cementing operation. In addition, the invasion
of filtrates into producing zones causes formation
damage, which can greatly reduce the production
potential of the reservoir. Developing cement
slurries that have minimal filtration loss prevents
expensive remedial cementing operations and reduces
formation damage. The OFITE Stirred Fluid Loss
Tester provides a reliable means of determining
the fluid loss characteristics of a well cement.
Method of Operation:
A cement slurry is poured into the test cell,
which is then placed into the heating jacket.
The gear drive system is connected to the agitation
paddle, which is dimensionally equivalent to an
atmospheric consistometer paddle. The desired
test temperature is maintained by a digital PID
temperature controller, while the necessary pressure
is applied to the cell to prevent evaporation
of the liquid phase. When conditioning the cement
in accordance to API Specification 10 guidelines,
the paddle is rotated at 150 RPM for 20 minutes.
Once the cement is conditioned, the test cell
is rotated 180 degrees and the desired differential
pressure is applied to the cell. The filtrate
is collected in a back pressure receiver for 30
minutes. The API defines fluid loss as the volume
(cc) of filtrate that is collected during this
30-minute interval.
Features and Specifications:
- Maximum pressure of 2,000 PSI
- Maximum temperature of 400°F
- Temperature is maintained by PID temperature
controller
- Variable paddle rotation speed (5 to 200 RPM)
- Filtration portion of cell is dimensionally
equivalent to an API approved HTHP test cell
Instrument Requirements:
- 115-Volt, 60 Hz, 18-Amp power source
- 230-Volt, 50 Hz, 10-Amp power source
Crated Size: 38" × 26" × 34" (97 × 66 × 86 cm)
Crated Weight: 220 lb (998 kg)
Components: -
120-70
STIRRED FLUID LOSS TESTER, 115 VAC
-
120-70-1
OFITE MODEL 40 STIRRED FLUID LOSS TESTER (220 VOLT)
For more information about this product or to request a quote, contact us.
|